Objective Gradually increasing number of minimally invasive spine surgery (MISS), there is an increasing risk of radiation exposure to medical personnel during the surgery. We measured the radiation exposure of the operating room personnel during MISS, tried to find the riskiest person, and checked the effectiveness of a new lead-composite shielding curtain.
Methods Radiation exposure of medical staffs (operator, first assistant, anesthesiologist, and scrub nurse) involved in MISS procedures of 35 patients without shielding curtain (nonshield group) and 35 patients with shielding curtain (shield group). The shielding curtain had 0.25-mm nominal lead equivalent and was mounted on 2 frame bars fixed on the operating table.
Results In the nonshield group, radiation exposure was significantly higher in the order of operator > first assistant > scrub nurse > anesthesiologist (p < 0.001) during both anteroposterior (AP) and lateral views. In the shield group, the radiation exposure of the operator and the scrub nurse decreased significantly by 94.1% and 76.4% in AP view (p < 0.001), and by 96.3% and 73.9% in lateral view (p < 0.001), respectively.
Conclusion Since the radiation dose of the operator was highest in a C-arm-guided MISS, there is a high priority need to protect the operator from the radiation exposure. The shielding curtain could most effectively reduce the radiation exposure of the operator.
Citations
Citations to this article as recorded by
Navigation and Robotic Single-Position Prone LLIF: First Cases in Brazil Rodrigo de Souza Lima, Luciano de Almeida Ferrer, Luciana Feitosa Ferrer, Vivian Nogueira de Castro Lima, Renata Silva Amaral World Neurosurgery.2024; 191: e72. CrossRef
Computer-Assisted Navigation Is Associated With Decreased Rates of Hardware-Related Revision After Instrumented Posterior Lumbar Fusion Patawut Bovonratwet, Alex Gu, Aaron Z. Chen, Andre M. Samuel, Avani S. Vaishnav, Evan D. Sheha, Catherine H. Gang, Sheeraz A. Qureshi Global Spine Journal.2023; 13(4): 1104. CrossRef
Unilateral Percutaneous Kyphoplasty Using a Novel Guide Device for Thoracolumbar Osteoporotic Vertebral Fracture Hongwei Wang, Hailong Yu, Yunpeng Zhu, Hongwen Gu, Bin Zheng, Yuanhang Zhao, Wenfeng Han, Liangbi Xiang Orthopaedic Surgery.2023; 15(4): 993. CrossRef
Radiation-Shielding Devices: The Best Combination for Spine Interventional Procedures Young Seok Ji, Eun Kyul Park, Hyuk Chan Kwon, Woong Ki Han, Francis Sahngun Nahm Journal of Vascular and Interventional Radiology.2022; 33(3): 225. CrossRef
Efficiency of novel shielding curtains combined with pulsed irradiation for reducing radiation exposure in an operating room: Human renal collecting system phantom study Yasuhiro Kaku, Takaaki Inoue, Yu Charlie, Satoshi Imai, Fukashi Yamamichi, Masato Fujisawa International Journal of Urology.2022; 29(6): 571. CrossRef
Effectiveness of a radiation protective device for anesthesiologists and transesophageal echocardiography operators in structural heart disease interventions Hitoshi Miyazaki, Yoshiyuki Umezu, Emi Kinoshita, Kazuhisa Ogawa, Ken-ichi Hiasa, Hiromichi Sonoda, Yuji Karashima, Toyoyuki Kato, Akira Shiose Cardiovascular Intervention and Therapeutics.2021; 36(4): 523. CrossRef